AL-204 Aluminum
Metal Analyzer
The AL-204 Aluminum Metal Analyzer is a compact, fully automatic device designed to assess the concentration of metallic aluminum in samples using gasometry. It performs fast tests with high precision and excellent reproducibility.
Accesories
| Reference | Description | Minimun quantity |
|---|---|---|
| AL-144 | Glass crucible for samples | 1 |
| CAC-202-81 | Glass crucible Kitasatos | 1 |
| EQFX-208 | Magnetic stirrer 30x6mm | 1 |
| AL-104 | O’Ring 30x3, Kitasatos sealing | 1 |
| AL-140 | Silicone gasket, transparent, piston sealing (sample crucible) | 1 |
Request more information
The AL-204 is a compact, fully automatic instrument designed to assess the concentration of metallic aluminum in a sample using gasometry. The analysis is carried out by continuously monitoring the pressure increase generated by the hydrogen released during the oxidation reaction of metallic aluminum:
2 Al (solid) + 2 NaOH + 6 H2O 🡪 2 Al(OH)4– + 2 Na+ + 3 H2 (gas)
The AL-204 Aluminum Analyzer is easy to operate. The sample is placed directly into a glass crucible, where the oxidation reaction occurs. Once the sample is loaded, the analyzer seals the system, dispenses the reagent, and continuously monitors the metallic aluminum concentration until the final result is obtained.
A user-friendly interface allows for sample data input, execution of the analysis, real-time graphical representation of the process, and automatic storage of results in a database, simplifying diagnostic and analytical tasks.
The AL-204 features two independent measurement channels, enabling either double the sampling speed or simultaneous analysis of different sample types.
Technical specifications
| Measurement Range: | 0 a 100% Al |
| Result Expression: | % de Aluminum purity |
| Sample Weight: | 0.5 g to 5 g; (máx. 0.5 g for Al standard) |
| Analysis Time: | Programmable 0 to 45 min (max 10 min for Al standard) |
| Max Analysis Pressure: | 1250 mmbar (Max detector pressure 1600 mbar) |
| Crucible Temperature: | Ambient to 80°C (Recommended: 40°C) |
| Accuracy: | <1.0% |
| Resolution: | 0.1% |
| Regents: | Aluminum standard (>99%) and sodium hidroxide (20% w/w) |
| Operating Temperature: | 10°C to 35°C |
| Power Supply: | 220 VAC, 50 /60 Hz, 600 W |
| Pneumatic Pressure: | Instrumental air max 4.0 kg/cm2 (60 psi) |
| Dimensions: | 470 (H) x 600 (W) x 570 mm (D) |
| Weight: | 20 kg |
More equipment of interest
No existen equipos relacionados.